FEATURES:
Ø Support all kinds of full battery, half battery, three-electrode test, suitable for battery material and finished product research, farad capacitor material and finished product research, 0~5V full range controllable.
Ø The channels support parallel connection with a maximum output of 160A.
Ø Supports precision research and testing with 0.01% voltage and current accuracy.
Ø Support high-rate fast charge research, 30C can achieve 0.01% capacity accuracy; 100C can reach 0.1% capacity accuracy.
Ø Support custom pulse test with a minimum pulse width of 2mS.
Ø Full coverage of the measurement records of the voltage, current, capacity, energy, efficiency, capacitance, internal resistance and other parameters required for the study.
Ø The device comes with UPS and storage to effectively prevent data loss and damage caused by downtime, sudden power failure, network failure, etc.
D350A_V5C40AT8 PERFORMANCE:
ITEM | INSTRUMENT PERFORMANCE |
RANGE | GEAR | RANGE | ACCURACY(SET,READ) | RESOLUTION(SET,READ) |
Voltage1 | 5V (DC to -5V)*1 | ±0.01%RD±0.01%FS | ≤50uV, ≤10uV |
Current1 | 10mA | ±0.01%RD±0.01%FS | ≤100nA, ≤20nA |
Current 2 | 100mA | ±0.01%RD±0.01%FS | ≤1uA, ≤200nA |
Current 3 | 1A | ±0.01%RD±0.01%FS | ≤10uA, ≤2uA |
Current 4 | 12A | ±0.01%RD±0.01%FS | ≤120uA, ≤24uA |
Current 5 | 40A | ±0.01%RD±0.01%FS | ≤400uA, ≤80uA |
Current set range | 1uA ~ 40A (MAX320A In parallel mode) |
No. of CHANNELs |
| 8Chl |
No. of STEPs |
| 1024 |
TIME | Accuracy | ≤±5uS/S (Equipment autonomous control, no communication and other time delay errors) |
Min Step Time | 20mS |
Min Pulse Time | 2mS |
Sampling Interval | ≤10mS*2 |
INPUT IMPEDENCE |
| ≥10GΩ(INPUT BASE CURRENT < 2.0nA) |
TEMPERATURE DRIFT |
| ≤20ppm/℃ |
RESPONSE TIME |
| Current 0~90% ≤1mS; Current -90~90% ≤3mS |
DATA RECORD | Sampling Rate | 100Hz(Sampling interval 10mS)*2 |
Setting Method | Interval time, Sampling rate, Interval current, Interval capacity, Interval voltage, Interval energy |
PROCESS CONTROL | Work Mode | HOLD / CC / CV / CCCV / CRATE / CP / CR / CPULSE / CSLOPE / DC / DV / DCDV / DRATE / DP / DR / DPULSE / DSLOPE |
End Conditions | TIME / VOLTAGE / CURRENT / POWER / CAPACITY / ENERGY / LOOPS / PHASE CYCLE |
Protective Function | Power-off data protection, Net break protection User-defined safety: Over voltage, Under voltage, Over current, Over rate, Over capacity. |
FUNTION |
| DCIR、Capacitance capacity calculation、Step pluse、Three-electrode test(Reference test)、DQ/DV |
OUTPUT |
| Four electrodes (support reference electrode test) |
COMMUNICATION |
| 10/100M Ethernet(RJ45)、USB(Console) |
OPERATING ENVIRONMENT | Temperature | 0~40℃(23±2℃Optimum accuracy) |
Humidity | ≤90% No condensation |
POWER SUPPLY |
| AC220V~±10% 50Hz / AC110V~±10% 60Hz, 2600W 3 |
EQUIPMENT SIZE |
| 6U W*H*L:430*540*178(mm) 32KG |
Note: 1. The voltage range can be customized by 2~8V.
2. The sampling rate can be customized to 1000Hz (sampling interval 1mS).
3. For best test accuracy and safety, be sure to ground well. The machine needs to be warmed for 30 minutes before use.
PACKING LIST:
No. | NAME & SPECIFICATION | QUANTITY | NOTE |
1 | D350A Device | 1 |
|
2 | Power Cable*1.5M | 1 | Multi-machine optional one-to-multiple power cord |
3 | Net Cable*2M | 1 |
|
4 | Certificate of Warranty | 1 |
|
5 | Factory Calibration Report | 1 |
|
6 | Test System Software (USB Disk) | 1 | System instructions is inside. |
| 1.2M Alligator Clamp |
|
|
1.2M Polymer Clamp |
3M Alligator Clamp |
3M Polymer Clamp |
Customized Clamp |
8 | Packing list | 1 |
|